Controller / User Interface | 4-37
CAL Mode – F5 Commands
Keystroke Term Denition
F5
F5 –
F5 0 FlowCtrlz The analyzer system Sample Flow Control mode, where ‘z = 0..2.
0 = Analyzer Control mode (automatic control by analyzer)
1 = Continuous Backpurge/Zero Flow mode (manually set by operator)
2 = Continuous Sample Flow mode (manually set by operator)
F5 1 Scalez The full-scale range for each of the four outputs, where ‘z = 1..4. If an output is not used,
its full-scale range should be set to ‘0’. For example, if Scale 1 = 1000 PPM, Output 1 would
read 20 mA at 1000 PPM and 4 mA at 0 PPM. These outputs are application-specic.
1 = Scale1
2 = Scale2
3 = Scale3
4 = Scale4
F5 2
F5 3
F5 4 CS2Dyn The existing CS
2
concentration (PPM) used for the dynamic compensation of the H
2
S and
SO
2
signals for CS
2
(applications with Standard Software version only – if measuring).
F5 5 CS2Histz The history of the last nine CS
2
readings, where ‘z = 1..9, with ‘1’ being the most recent
value (other than the current value) and ‘9 being the oldest.
These values are updated 15 minutes after every Auto-Zero, which makes it possible
to look back at the CS
2
readings for a period of just over two hours. However, updating
only occurs if the analyzer CS
2
compensation is dynamic (applications with Standard
Software version only – if measuring).
F5 6 T0z ** The measuring wavelength transmittance value with Zero gas for each lter, where
z = 1..6. Adjusted automatically whenever the analyzer is Zeroed.
1 = T01 4 = T04
2 = T02 5 = T05
3 = T03 6 = T06
F5 7 SFactorz The Span (calibration) factor for the calculated result (component concentration), where
z = 1..2. Adjusted automatically whenever the analyzer is Spanned.
1 = SFactor1
2 = SFactor2
F5 8 Temp The default temperature for the Measuring Cell (°C or °F). This value is used in place of the
measured Measuring Cell temperature when active temperature compensation is disabled.
F5 9 Pres The default absolute pressure for the Measuring Cell (mmHg or "Hg). This value is used
in place of the measured Measuring Cell pressure when active pressure compensation is
disabled.