EasyManuals Logo

Ametek 930 User Manual

Ametek 930
256 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #32 background imageLoading...
Page #32 background image
2-4 | Model 900 ADA / Model 930 Sulfur Pit Analyzers
24-Hour Zero Drift
SO
2
: 0.5 % of full-scale of standard ranges
H
2
S: 1 % of full-scale of standard ranges
Model 900 only:
COS, CS
2
: 10 % of full-scale of standard ranges
Customer Connections
Analog Outputs
4 isolated 4–20 mA self-powered (or optional loop-powered; loop power
by end user).
Digital Communication
One RS-232 port used for service diagnostics.
One RS-422 port for remote communication with a Data Acquisition
System (DAS), using Modicon Modbus
®
protocol.
Optional RS-485 for Modicon Modbus
®
protocol (requires RS-232 to
RS-485 converter).
Optional System 200 Configurator Software.
Maximum Sample Gas Pressure
Maximum Sample
Pressure Transducer Gas Pressure
100 psia 350 kPag
30 psia 63 kPag
Electrical Requirements
Maximum Start-Up Power
<750 W maximum start-up with continuous average, depending on
ambient temperature (includes 150 W for ASR900 Sample Probe; excludes
Sample and Vent Lines).
Supply Voltage
120 VAC (± 10 %), 47–63 Hz
240 VAC (± 10 %), 47–63 Hz

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek 930 and is the answer not in the manual?

Ametek 930 Specifications

General IconGeneral
BrandAmetek
Model930
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals