Maintenance and Troubleshooting | 5-73
Analyzer Reset
If the Host Controller board or Microcontroller board must be reset, take
all necessary safety precautions and then simultaneously press and Ent
on the User Interface. The message “Reset In Progress” should be dis-
played within one second after pressing these keys.
Do not randomly reset the analyzer during normal operation.
Typically, the only resets required are in cases where the Host
Controller or Microcontroller boards have been replaced. In this
case, it is necessary to shut down the analyzer. The analyzer will reset
upon normal power-up.
!
CAUTION