4-52 | Model 900 ADA / Model 930 Sulfur Pit Analyzers
Example 1:
To calibrate the current output signal for Output Channel 1 and return to
CAL mode normal display, press:
F6 0 1 Del Sets the output to zero-scale.
3.97 Ent The current meter reading is 3.97 mA.
F6 1 1 Del Sets the output to full-scale.
19.92 Ent The current meter reading is 19.92 mA.
Ent Returns the CAL mode normal display.
Example 2:
To calibrate the current output signal for Output Channel 2 and return to
CAL mode normal display, press:
F6 0 2 Del Sets the output to zero-scale.
4.06 Ent The current meter reading is 4.06 mA.
F6 1 2 Del Sets the output to full-scale.
20.18 Ent The current meter reading is 20.18 mA.
Ent Returns the CAL mode normal display.