User Manual Rev BE AMETEK Programmable Power
MX Series 130
REPEAT This is the number of times the FREQUENCY
SWEEP/STEP transient will repeat before it proceeds to
the next event or exit the transient. The number of
times the transient event is generated is equal to the
REPEAT + 1. Leave this value at zero if only one
execution of this event in the list is required.
EVENT # This must be the last item in the transient edit screen.
All data fields must be entered before inserting the
EVENT #. The EVENT # takes value from 1 to 99. The
EVENT # defines the order of execution of the transient
events in a multiple event transient. It is a good
practice to enter spaced EVENT #’s to allow insertion of
an EVENT later if needed. (For example, space them by
5.) Entry of a sequence EVENT # number will cause
the display to return to the TRANSIENT screen.