User Manual – Rev BE    AMETEK Programmable Power 
MX Series  232 
selected or use the PHASE key if not. Do the same with the IMP REACT FS field. 
Note that the adjustment range for R is 0 to 100, for L is 0 to 300.  
Press SET to save the calibration coefficients for the selected PHASE. 
If there is not enough range in the full-scale calibration coefficient for either 
resistive or inductive portion, it may be necessary to tweak the adjustment pots 
on the MX controller.  These pots were originally adjusted at the factory and 
normally do not have to be adjusted again. The Full-Scale calibration 
coefficients should have enough adjustment range.  Double check the 
connections and phase measurements if this is not the case to make sure the 
measurement readings you get are indeed correct. 
If it is necessary to adjust the pots, see Table 6-6 for the corresponding pot 
designators. The top cover must be removed to access these pots. They are 
located along the top edge of the controller board(s).  For Series I MX units, 
adjustments for phase A are on the phase A board (7000-721), adjustments for 
phase B and C are on the phase B/C board (7000-722). For Series II MX units, 
all adjustments are on the 7003-718 controller board. The reference designators 
are the same on either series controller. 
Repeat steps 2 through 11 for phase B and C. (except on single phase only 
models). 
Phase A  (7000-721) 
R121  R122 
Phase B  (7000-722) 
R112  R111 
Phase C  (7000-722) 
R114  R115 
Phase A  (7003-718-2 /  -4) 
R121  R122 
Phase B  (7003-718-2) 
R112  R111 
Phase C  (7003-718-2) 
R114  R115 
Table 6-6: Programmable Z Adjustment Pots