4.2.13 MEASUREMENT CAL FACTORS Menu
Figure 4-28: MEASUREMENT CAL FACTORS Menu (Series I only)
Figure 4-29: MEASUREMENT CAL FACTORS Menu (Series II only)
The MEASUREMENT CAL FACTORS menu provides access to the measurement
calibration parameters. The parameters apply to the selected mode of operation
(AC or DC mode). For three phase configurations, the PHASE keys toggle
between the three calibration screens for each phase. These parameters are
password protected and can only be changed after the calibration password has
been entered. Refer to the calibration section in this manual for details on
performing a calibration.
Note that Series II MX models only have two measurement calibration factors.
Series I models require additional cal factors as shown. The following calibration
factors are available from this menu:
Entry Description
VOLT FS Full scale voltage measurement calibration factor.
VOLT 0 Zero offset voltage measurement calibration factor.
(Series I only)
VOLT FS HT Temperature compensated full-scale voltage
measurement calibration factor. (Series I only)
VOLT 0 HT Temperature compensated zero offset voltage
measurement calibration factor. (Series I only)
CURR FS Full scale current measurement calibration factor.
CURR 0 Zero offset current measurement calibration factor.
(Series I only)