4.7.8  Transient Execution 
 
Figure 4-46: START/VIEW TRANSIENT SEQUENCE Menu 
A transient list can be executed from the START/VIEW TRANSIENT SEQUENCE 
menu. To start a transient list, position the cursor on the START field as shown 
in Figure 4-46 and press the ENTER key. Transients may be aborted by pressing 
the ENTER key again while on the same field as the field changes to ABORT 
while a transient execution is in progress. For short duration transients, this will 
likely not be visible, as the transient will complete before the screen is updated. 
Longer duration transients however may be aborted in this fashion. 
Longer duration transients may also be suspended using the PAUSE field located 
below the START/ABORT field. Pressing the ENTER key while on the PAUSE field 
will suspend the transient execution. Once suspended, it can be resumed using 
the same field as the field changes to RESUME while the transient execution is 
suspended. Suspending a transient may be useful when running slowly changing 
output transients to ‘hold’ the output at a specific setting while observing the 
effect on the unit under test.