Figure 9-25: IEC 1000-4-13 Harmonics 
LEVEL 
Sets the percentage level of the harmonic relative to the fundamental. Each 
level is associated with a harmonic number. To change levels, the test state 
must be off and user class must be selected. 
HARM NO 
Selecting a harmonic number using the shuttle knob will allow examination of 
the LEVEL and PHASE of each harmonic and changing the level and phase angle 
when USER class is selected. 
PHASE 
Sets the phase angle of the harmonics relative to the fundamental. To change 
phase, the test state must be off and user class must be selected. 
PAUSE 
Sets the pause time in seconds between harmonics tests. There is only one 
value for the entire test. The PAUSE value can be changed when the test is not 
running. 
INTERHARMONICS GROUP 
If the Harmonics group is selected, the screen will appear as shown in Figure 
9-26.  The screen has the following parameters that are unique to the group: 
 
Figure 9-26: IEC 1000-4-13 INTERHARMONICS 
LEVEL 
Sets the percentage level of the interharmonics relative to the fundamental. The 
level is fixed for the entire frequency range defined by FRANGE. To change 
levels, the test state must be off and user class must be selected. 
FRANGE