User Manual – Rev BE AMETEK Programmable Power
MX Series 283
EUT CLASS
This field selects the desired product or EUT class. There are three EUT classes.
Class 2 and 3 are predefined by the standard and its level parameters cannot be
changed. The user class can be edited at any time. Changing between class 2
and 3 can be done while the state is on. Changing to the user class requires the
state to be in the off position.
For Class 1 EUT, the user class must be used. Only class 2 and 3 test levels are
pre-programmed. When using the provided, Virtual Panels Windows program,
class 1 tests are set up by the Virtual Panels program automatically.
GROUP
This field selects the GROUPs. Selecting a group is required for editing the data
fields associated with the group as well as running the individual group if run
group is selected. The field has the following groups:
FCURVE
This is the first part of harmonic combination test.
OSWING
This is the second part of harmonic combination test.
SWEEP
This is the sweep in frequency and resonance frequency detection.
HRAM
This is the first part of the Individual harmonics and interharmonics test.
IHARM
This is the second part of the Individual harmonics and interharmonics test.
MCURVE
This is the Meister test. (Firmware revision 0.31 or higher required)
See the appropriate sections for details on each group.
DWELL
This field will show the dwell time in seconds for each group during the test and
will allow changes to the dwell time to be made.