Page 10
Indications for Use
The Animas 2020 insulin infusion pump is intended to provide subcutaneous delivery of
insulin at programmable basal and bolus rates for the daily management of diabetes mellitus
in insulin dependent patients.
Contraindications
The pump is not indicated for anyone unable or unwilling to:
Test BG levels at least 4 – 6 times per day or as recommended by healthcare team
Demonstrate adequate carbohydrate counting skills
Maintain good diabetes self-care skills
See his/her healthcare team regularly
Adequate vision or hearing is required to recognize the pump alerts.
Warnings and Precautions
Radiology Equipment
Your Animas 2020 has been designed to operate in the presence of common sources of
electrostatic and electromagnetic interference, including airport and store security systems.
However, like all portable electronic devices, the pump should not be exposed to very
strong electromagnetic fields, such as in MRI’s, RF welders or magnets used to lift
automobiles. Very strong magnetic fields, such as in an MRI, can re-magnetize the portion
of the motor that regulates insulin delivery. If you plan to undergo an MRI, remove your
pump beforehand and keep it outside the room during the procedure.
Medical Procedures
Procedure Patient is Pumper Healthcare Professional/Technician is
Pumper
Pacemaker / AICD
Disconnect pump and leave outside room
during insertion of device and reprogramming.
Infusion set can remain in place.
Follow normal safety practices and proceed
to the designated safe area while each X-
ray is taken and during reprogramming.
EKG
No need to disconnect. No need to disconnect.
Cardiac Catheter
Disconnect pump and leave outside room
during procedure.
Follow normal safety practices and proceed
to the designated safe area while each X-
ray is taken.
Nuclear Stress test
Disconnect pump and leave outside room
during scan. Can remain connected during
injection of radioisotope.
Follow normal safety practices and proceed
to the designated safe area while each X-
ray is taken.
Colonoscopy
No need to disconnect. No need to disconnect.
Laser Surgery
The pump and infusion set can be worn;
however some lasers can create interference
and cause the pump to alarm.
The pump and infusion set can be worn;
however some lasers can create
interference and cause the pump to alarm.
General anesthesia
Determination based on what medical
equipment is being used in the procedure.
Determination based on what medical
equipment is being used in the procedure.
For procedures not included above or on the chart on the following page, call (in advance
of your procedure) 1-877-767-7373 in the US, Canada and Puerto Rico. Outside the US,
Canada and Puerto Rico, contact your local Animas distributor.