Page 81
Chapter 14 Safety System and Alarms
Alerts, Warnings and Alarms
Your Animas pump has a progressive Warnings and Alarms safety system. This means if you
do not confirm the Warning or Alarm, it will progress to the sweep alarm with vibrate
within one hour. At the high volume stage, if you do not confirm the Warning or Alarm,
the sweep alarm will begin and will not stop until appropriate action is taken.
Note: Your pump uses battery power to notify you of reminders, alerts, warnings and
alarms. If you do not confirm the notification, your pump will continue to use battery
power as the notifications repeat and progress. This will result in reduced battery life and
the Replace Battery Alarm screen appearing sooner than expected.
Additionally, certain warnings (e.g., Empty Cartridge Alarm) take precedence over less
critical ones (e.g., Low Battery Warning). This means if you do not confirm the more
critical warning, battery life will be reduced and your pump may skip the Low Battery
Warning and go directly to the Replace Battery Alarm, or the battery life will end before a
Replace Battery Alarm is displayed.
Alerts are automatically displayed to remind you of a function that you’ve set or a condition
that exists. Warnings are triggered for a variety of reasons. They require you to confirm
the warning by pressing OK and/or taking action to address the warning.
Alarms are triggered by several conditions. All require you to address the alarm by taking
appropriate action in order to clear the alarm condition.
- Indicates that this Alert, Warning or Alarm can play a tune as the initial notification.
Refer to your ezManager User Guide included with the software.
Alert: Active Basal Program Empty
Cause Empty basal program activated.
Effect No basal deliveries.
Message Displayed once for 3 seconds and each time manually
awakened.
Action None required but can confirm or select Basal Menu.
Beeps/Vib User selected, one time and each time manually
awakened. No progression.
Alert: Temp Basal Minimum Rate
Cause Negative Temp Basal activated.
Effect Basal delivery will not go below 0.025 U/hr.
Message Displayed once for 3 seconds.
Action None required.
Beeps/Vib User selected, one time. No progression.