Page 82
Alerts, Warnings and Alarms
Alert: Low BG
Cause BG entry below 70 mg/dL.
Effect Requires user confirmation to continue.
Message Displayed until confirmed.
Action Press OK to confirm.
Beeps/Vib
User selected, one time and each time manually
awakened until confirmed. No progression.
Alert: High BG
Cause BG entry above 250 mg/dL.
Effect Requires user confirmation to continue.
Message Displayed until confirmed.
Action Press OK to confirm.
Beeps/Vib
User selected, one time and each time manually
awakened until confirmed. No progression.
Alert: Clear Program Basal Segments
Cause Clear command selected from Basal Options screen.
Effect Requires user confirmation to continue.
Message
Displayed until confirmed or until pump goes to
sleep.
Action Select Clear Program or Basal Options.
Beeps/Vib User selected, one time.
Alert: Suspend (Temp Basal/Combo Bolus Canceled)
Cause Pump suspended.
Effect Any active Temp Basal/Combo Bolus canceled.
Message Displayed once for 3 seconds.
Action None required.
Beeps/Vib User selected, one time. No progression.
14