Amusement Parks
Very powerful electromagnets are sometimes used on “free-fall” amusement park rides.
Insulin pumps should be removed and not taken on these “free-fall” types of rides.
High gravity forces can be experienced when riding on some roller-coasters. It is
recommended that you disconnect (NOT suspend) the pump while on roller-coaster rides.
Aircraft without Cabin Pressurization
If flying in aircraft without cabin pressurization or flying in aircraft used for aerobatics
or combat simulation (pressurized or not), it is recommended that you disconnect (NOT
suspend) the pump.
Precautions
The Animas pump is a sealed device that should be opened ONLY by the manufacturer.
If the pump seal is broken by anyone other than an authorized Animas factory
technician, the warranty is voided and the pump is no longer waterproof. If the back
label on the pump is removed or damaged, the warranty is voided and the pump is no
longer waterproof.
When using an insulin pump, if the pump is placed at a vertically higher position than
the infusion site, a very small amount of additional insulin infusion may occur. To
minimize this condition and maintain pump delivery accuracy, the vertical distance
between the pump and the infusion site should be no more than 12 inches (30 cm). If
the pump is placed at a vertically lower position than the infusion site, this condition is
eliminated.
Occasionally check the infusion site for proper placement and leaks. Improperly placed
infusion sites or leaks around the infusion site can result in under infusion.
Occasionally check the infusion set tubing for any damage, leaks or kinks while using
the pump. Damaged, leaking or kinked tubing may restrict or stop insulin delivery and
result in under infusion.
Only use Animas 1200/2000 series (200 unit/2ml) cartridges and infusion sets with a
standard Luer™ connector. Efficacy of pump cannot be guaranteed if cartridges other
than those manufactured by Animas Corporation are used or if cartridges are used more
than once.
Always dispose of used cartridges and infusion sets following the regulations in your
community. Insulin can be absorbed through the skin. Minimize any contact with
insulin and wash hands thoroughly after handling any devices that contain insulin.
Failure to follow these guidelines may pose health hazards.
Prior to replacing the battery cap, make sure the o-ring (yellow) fits securely and is not
damaged. See Chapter 5, Getting Your Pump Ready.
Prior to inserting a cartridge into the pump, check the o-rings on the cartridge to be
sure they are not damaged.
Occasionally check the vents to be sure they are clear of debris. See Chapter 13, Care and
Maintenance.