If you drop the pump or it has been hit against something hard, inspect it to be sure it
is still working properly. Check that the display screen is working and clear, that the
cartridge cap, battery cap and infusion set are properly in place. Check for leaks around
the cartridge by wrapping a piece of tissue around the connection area. Cracks, chips or
damage to the pump may impact the battery contact and/or the waterproof feature of
your pump. Call our 24/7 Pump Support Department at 877-767-7373 if you identify or
suspect your pump has been damaged. They will help determine if your pump should be
replaced. (Outside the US, Canada and Puerto Rico, contact your local Animas
distributor.)
The pump is designed to operate in conditions where temperatures are between 40º and
104ºF (5 – 40C). If your pump is exposed to temperatures outside these parameters,
extra care should be taken to protect it from extreme temperatures.
Animas insulin pumps and insulin pump cartridges are latex free.
To avoid risk of explosion, do not use the pump in the presence of explosive gases.
Your 2020 pump is designed to achieve optimum performance and battery life with an
Energizer Lithium L91 AA battery (1.5V). Rechargeable batteries and Carbon-Zinc
batteries do not have the necessary characteristics to power the pump and should not be
used. Some AA lithium batteries are available with other voltages such as 3.6V or 4.0V.
Do not use these batteries. Use of anything other than a 1.5V battery could
permanently damage the pump and voids its warranty.
Do not use household cleaners, chemicals, solvents, bleach, scouring pads or sharp
instruments to clean the pump. Never put your pump in the dishwasher or use very hot
water to clean it.
Never use a hair dryer, microwave oven or baking oven to dry your pump. Use a soft
towel.