Alerts, Warnings and Alarms
Warning: Low Cartridge
Cause Low insulin level reached.
Effect
Deliveries may continue until Empty Cartridge alarm
is triggered.
Message Displayed when manually awakened until confirmed.
Action Press OK to confirm. Replace with filled cartridge.
Beeps/Vib
User selected, every 3 minutes until confirmed. If not
confirmed, progresses to sweep/vibe within one hour.
Warning: Exceeds Max Bolus
Cause Audio bolus delivery exceeds user-set maximum.
Effect Bolus delivery stops.
Message Displayed when manually awakened until confirmed.
Action
Press OK to confirm. Reprogram max bolus amount in
Setup Advanced.
Beeps/Vib
User selected, every 3 minutes until confirmed. If not
confirmed, progresses to sweep/vibe within one hour.
Warning: Exceeds Max TDD
Cause Bolus delivery exceeds user-set maximum.
Effect All deliveries may stop until action is taken.
Message Displayed when manually awakened until confirmed.
Action
Press OK to confirm. Reprogram max TDD amount in
Setup Advanced.
Beeps/Vib
User selected, every 3 minutes until confirmed. If not
confirmed, progresses to sweep/vibe within one hour.
Warning: Low Battery
Cause Battery life ends in minimum 1 hour.
Effect Deliveries continue.
Message
Displays when pump is awake until confirmed.
Displays when triggered by event (such as bolus) &
when manually awakened.
Action Press OK to confirm. Insert new battery.
Beeps/Vib
User selected, every 3 minutes until confirmed. If not
confirmed, progresses to sweep/vibe within one hour.