Page 86
Alerts, Warnings and Alarms
Alarm: Empty Cartridge
Cause Cartridge empty.
Effect All deliveries stop.
Message Continuous until confirmed.
Action
Press OK to confirm. Replace with full cartridge.
Option to select Suspend (see Suspend).
Beeps/Vib
User-selected, every 3 minutes. If not confirmed,
progresses to sweep/vibe within one hour. (Once
confirmed, No Prime warning triggered, see No Prime.)
Alarm: Replace Battery
Cause
Battery either has minimum 3 minutes left or battery
is dead.
Effect All deliveries stop.
Message Continuous until battery is removed.
Action Remove battery to silence alarm. Insert new battery.
Beeps/Vib MAX volume every 3 minutes until action taken.
Alarm: Call Service
Cause Hardware or software problem detected.
Effect All deliveries stop.
Message Continuous until battery is removed.
Action
Press OK to silence alarm for 30 minutes (alarm can
only be silenced once). Call Pump Support. (Outside
the US & Canada, contact your local Animas
distributor.)
Beeps/Vib
User-selected, every 3 minutes. If not confirmed,
progresses to sweep/vibe within one hour.
Alarm: Auto-Off
Cause Suspended due to no key press in user-set time period.
Effect All deliveries stop.
Message Continuous until confirmed.
Action
Press OK to confirm. Adjust time period in Setup
Advanced.
Beeps/Vib
MAX volume every 3 minutes until action taken.
(Once confirmed, No Prime warning triggered, see No
Prime.)
14