Instrument Notifications and Troubleshooting A-7 System Recovery
MS2090A UG PN: 10580-00444 Rev. U A-23
Install SW to SLOT X
The recovery mode shows two slots (A and B). These slots contain different system software packages that are
installed. Each slot contains the following information:
Slot name (A,B): Boot version package (A or B).
Slot count (3,2,1): The count indicates how many unsuccessful boot attempts before the slot is marked as
unbootable. Every failed attempt decrements the value until it reaches 0 (unbootable). This count can be
reset to 3 using the “Reset” menu.
Slot version: The version represents the software package version available in the slot.
You can install a new software package by inserting a USB memory device with the system software file
(.raucb) and pressing “Install SW to SLOT X.” The X will be replaced with an A or B depending on which slot is
to be overwritten. You will not be able to select which system software file will be installed from the attached
USB, it will always install the first one found.