EasyManua.ls Logo

Anritsu LMR Master S412E - Page 105

Anritsu LMR Master S412E
704 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Spectrum Analyzer Commands 4-14 [:SENSe] Subsystem
S412E PM PN: 10580-00319 Rev. N 4-63
[:SENSe]:OBWidth:METHod XDB|PERCent
[:SENSe]:OBWidth:METHod?
Title: Occupied Bandwidth Measurement Method
Description: Sets the method for calculating occupied bandwidth. XDB calculates the
occupied bandwidth based on points a specified number of dB below the
carrier. Issue command [:SENSe]:OBWidth:XDB to set the number of
dB to be used. PERCent calculates the occupied bandwidth based on
points a specified percentage of the carrier power below the carrier.
Issue command [:SENSe]:OBWidth:PERCent to set the percentage to be
used.
Parameter: XDB|PERCent
Parameter Type: <char>
Default Value: PERCent
Related Command: :OBWidth:XDB :OBWidth:PERCent
Front Panel Access: Shift-4 (Measure), OCC BW
[:SENSe]:OBWidth:PERCent <percentage>
[:SENSe]:OBWidth:PERCent?
Title: Occupied Bandwidth Percent of Power
Description: This command sets the percentage of carrier power used to measure the
occupied bandwidth. This value is used in the measurement if
:SENSe:OBWidth:METHod is set to PERCent.
Parameter: <percentage>
Default Value: 99
Default Unit: %
Range: 0% to 100%
Related Command: :OBWidth:METHod
Front Panel Access: Shift-4 (Measure), OCC BW, %
[:SENSe]:OBWidth:STATe OFF|ON|0|1
[:SENSe]:OBWidth:STATe?
Title: Occupied Bandwidth Measurement State
Description: Sets the state of the occupied bandwidth measurement, ON or OFF.
When using :CONFigure:OBWidth, the state is automatically set to ON.
Parameter: OFF|ON|0|1
Parameter Type: <boolean>
Default Value: OFF
Related Command: :CONFigure:OBWidth
Front Panel Access: Shift-4 (Measure), OCC BW, On/Off

Table of Contents