VNA Commands 5-3 :Display Subsystem
S412E PM PN: 10580-00319 Rev. N 5-35
:DISPlay:WINDow:TRACe:Y[:SCALe]:GDAPerture <integer>
:DISPlay:WINDow:TRACe:Y[:SCALe]:GDAPerture?
Description: Sets the Group Delay aperture value (which is common to all traces).
The query version of this command produces the Group Delay aperture
as its output.
Syntax: :DISPlay:WINDow:TRACe:Y[:SCALe]:GDAPerture <integer>
:DISPlay:WINDow:TRACe:Y[:SCALe]:GDAPerture?
Cmd Parameter: <NR1> <integer>
Query Response: <NR1> <integer>
Default Value: 2
Range: 2to20
Front Panel Access: Scale, Aperture
Note
Graph type must be Group Delay in order to display the Aperture % soft key in the
Scale menu.