5-11 [:SENSe] Subsystem VNA Commands
5-74 PN: 10580-00319 Rev. N S412E PM
[:SENSe]:CORRection:COLLect:METHod SOLT
[:SENSe]:CORRection:COLLect:METHod?
Description: Sets the calibration method.
Parameter: SOLT
Cmd Parameter: <char> SOLT
Query Response: <char> SOLT
Default Value: SOLT
Front Panel Access: Shift-2 (Calibrate)
[:SENSe]:CORRection:COLLect:SAVe
Description: Calculates the calibration data according to the calibration method that
is selected and the steps that are performed and then stores the result.
This command is similar to the “calculate and finish” on the front panel
of the list of calibration steps.
Cmd Parameter: NA
Query Response: NA (no query)
Front Panel Access: NA
[:SENSe]:CORRection:COLLect:STATus?
Description: This command requests information about the calibration status. The
command returns 0 if none, 1 if calibration has already started, 2 if
calibration has been aborted, 3 if a calibration is currently calculating,
and 4 if a calibration has been completed.
Cmd Parameter: NA (query only)
Query Response: <char> [<cal steps>, <port_no>]
Front Panel Access: NA