VNA Commands 5-11 [:SENSe] Subsystem
S412E PM PN: 10580-00319 Rev. N 5-75
[:SENSe]:CORRection:COLLect:STATus:ACCuracy?
Description: This command requests information about the calibration accuracy
status. The command returns:
0 when no calibration is available (CAL off)
1 when the calibration accuracy is high (OK: Accuracy High)
2 when the accuracy is moderate due to a change in power level
(?P: Accuracy Moderate)
3 when the accuracy is moderate due to a change in temperature
level by more than 5 deg C (?T: Accuracy Moderate)
4 when the accuracy is low due to a change in temperature level by
more than 10 deg C (X: Accuracy Low).
Cmd Parameter: NA (query only)
Query Response: <NR1> (integer)
0 = CAL off
1 = OK: Accuracy High
2 = ?P: Accuracy Moderate
3 = ?T: Accuracy Moderate
4 = X: Accuracy Low
Front Panel Access: Shift 2 (Calibrate), Existing Cal Info
[:SENSe]:CORRection:COLLect:TYPE <cal type>
[:SENSe]:CORRection:COLLect:TYPE?
Description: Configures the calibration type. <cal type> must be one of the
following values:
RFP1|TRFP|RRP1|2PFP
Syntax: :CORRection:COLLect:TYPE <cal type>
:CORRection:COLLect:TYPE?
Cmd Parameter: <char> <cal type>
Query Response: <char> <cal type>
Default Value: 2PFP
Front Panel Access: Shift-2 (Calibrate), Cal Type
CAL TYPE DESCRIPTION
RFP1 Full Port 1
TRFP Trans Response Fwd Path
RRP1 Reflection Response Port 1
2PFP 1 Path 2 Port Fwd Path