7-9 :READ Subsystem Fixed WiMAX Commands
7-20 PN: 10580-00319 Rev. N S412E PM
:READ:DEMod:EVSYmbol?
Title: Read EVM vs. Symbol
Description: Triggers a new EVM vs. Symbol measurement and returns the results:
EVM (rms), EVM (pk), RCE (rms), RCE (pk), Carrier Frequency, Freq
Error, Freq Error, and the Base Station ID The EVM vs. Symbol
measurement must be the active measurement (specified by
:CONFigure:DEMod EVSYmbol). The current measurement can be
queried using the command :CONFigure?
Data is returned as 8 comma-delimited values: EVM (rms) in %, EVM
(pk) in %, RCE (rms) in dB, RCE (pk) in dB, Carrier Frequency in MHz,
Freq Error in MHz, Freq Error in ppm, and the Base Station ID.
Related Command: :FETCh:DEMod:EVSYmbol?
:CONFigure:DEMod EVSYmbol
:READ:DEMod:SFLatness?
Title: Read Spectral Flatness
Description: Triggers a new Spectral Flatness measurement and returns the
absolute delta of the power between adjacent sub carriers in dB. The
Spectral Flatness measurement must be the active measurement
(specified by :CONFigure:DEMod SFLatness). The current
measurement can be queried using the command :CONFigure?
Related Command: :FETCh:DEMod:SFLatness?
:CONFigure:DEMod SFLatness
:READ:PFail?
Title: Read PASS/FAIL
Description: Triggers a new PASS/FAIL measurement and returns the results. It is a
combination of the commands :ABORT; :INITiate; :FETCh:PFail? The
PASS/FAIL measurement must be active. The current measurement
can be queried using :CONFigure?
Related Command: :FETCh:PFail?
:CONFigure PFail
:MEASure:PFail?
Front Panel Access: Measurements, Pass/Faile Mode
:READ:RF:ACPR?
Title: Read Adjacent Channel Power Ratio
Description: Triggers a new Adjacent Channel Power Ratio measurement and
returns the results: Power levels for each channel (both absolute and
relative). It is a combination of the commands :ABORT; :INITiate;
:FETCh:RF:ACPR? The ACPR measurement must be the active
measurement (specified by :CONFigure:RF ACPR). The current
measurement can be queried using :CONFigure?