P25 Phase 1 Commands 10-2 :CONFigure Subsystem
S412E PM PN: 10580-00319 Rev. N 10-3
:CONFigure:BITCap
Description: This command configures the P25 Bit Capture measurement. Certain
settings from the previous measurement (Ref Lvl, Scale, Graph Types,
Maximize) will be backed up when exiting this measurement and
restored when entering this measurement. All other settings carry over
from one measurement to the other. Please note that you must have the
Rx pattern set to VOICe to set the P25 Bit Capture measurement.
Cmd Parameters: NA
Query Parameters: NA
Range: NA
Default Value: NA
Default Unit: NA
Example: To switch the measurement to P25 Bit Capture:
:CONFigure:BITCap
Related Command: :DM:PATTern VOICe
Front Panel Access: Measurement, P25 Bit Capture
:CONFigure:CONTrol
Description: This command configures the P25 Control measurement. Certain
settings from the previous measurement (Ref Lvl, Scale, Graph Types,
Maximize) will be backed up when exiting this measurement and
restored when entering this measurement. All other settings carry over
from one measurement to the other. Please note that you must have the
Rx pattern set to VOICe or CTRLchan to set the P25 Control
measurement.
Cmd Parameters: NA
Query Parameters: NA
Range: NA
Default Value: NA
Default Unit: NA
Example: To switch the measurement to P25 Control:
:CONFigure:CONTrol
Related Command: :DM:PATTern VOICe
:DM:PATTern CTRLchan
Front Panel Access: Measurement, P25 Control