11-13 [:SENSe] Subsystem P25 Phase 2 Commands
11-40 PN: 10580-00319 Rev. N S412E PM
[:SENSe]:AVERage:COUNt <integer>
[:SENSe]:AVERage:COUNt?
Description: Sets the number of times the numerical values in the P25p2 Analyzer
Summary window are averaged.
Cmd Parameters: <integer>
Query Parameters: NA
Range: 1 to 25
Default Value: 1
Default Unit: NA
Example: To set averaging to 15:
:AVERage:COUNt 15
Front Panel Access: Setup, Averaging
[:SENSe]:CORRection:OFFSet[:MAGNitude] <value>
[:SENSe]:CORRection:OFFSet[:MAGNitude]?
Description: Sets the receiver power offset. Please note that when Auto Rx Range is
set to On, changing the offset value will cause the Ref Level to change.
For example, if the reference level is at 7.0 dBm and the Rx power offset
is then set to 10 dB external gain, the value of the refererence level will
be automatically adjusted down to -3.0 dBm.
If Auto Rx Range is Off, any adjustments to the offset will be reflected
in the vertical position of the spectrum trace. The reference level will
not be adjusted.
The Received Pwr value in the Summary Table is also affected by
changing this value.
The query returns a value from -100 to 100. A negative sign means
external gain and no sign means external loss.
Cmd Parameters: <value>
Query Parameters: NA
Range: 100 dB Ext Loss to 100 dB Ext Gain
Default Value: 0 dB
Default Unit: dB
Example: To set the external attenuation to 30 dB:
:CORRection:OFFSet 30
Front Panel Access: Amplitude, Rx Power Offset