11-13 [:SENSe] Subsystem P25 Phase 2 Commands
11-42 PN: 10580-00319 Rev. N S412E PM
The set command is sent using the units selected with the Rx Units
button on the front panel or with the command: UNIT:POWer:RX. If the
Rx Units has been set to dBm, the squelch setting must be set in dBm.
If the unit is set to watts, the squelch setting must be set in fW. If the
unit is set to Volts, the squelch setting must be set in fV.
Cmd Parameters: <value>
Query Parameters: NA
Range: 0 dBm to -120 dBm or 1 fW to 1000000000000 fW
or 223.6 mV to 223.61 nV
Default Value: -100 dBm or 100 fW or 2.24 μV
Default Unit: dBm or fW or fV
Example: To set the squelch to -10 dBm:
:DM:SQUelch -10
Front Panel Access: Setup, Squelch Lvl
[:SENSe]:FREQuency:CENTer <value>
[:SENSe]:FREQuency:CENTer?
Description: Sets the receiver center frequency. Please note that setting the center
frequency will restart the sweep
Cmd Parameters: <value>
Query Parameters: NA
Range: For 1.6 GHz Model: 100000 Hz to 1600000000 Hz
For 6 GHz Model: 100000 Hz to 6000000000 Hz
Default Value: 800000000 Hz
Default Unit: Hz
Example: To set the center frequency to 145 MHz:
:FREQuency:CENTer 145000000
Front Panel Access: Frequency, Rx Freq