dPMR Commands 13-13 [:SENSe] Subsystem
S412E PM PN: 10580-00319 Rev. N 13-37
[:SENSe]:FREQuency:CENTer <value>
[:SENSe]:FREQuency:CENTer?
Description: Sets the receiver center frequency. Please note that setting the center
frequency will restart the sweep
Cmd Parameters: <value>
Query Parameters: NA
Range: For 1.6 GHz Model: 100000 Hz to 1600000000 Hz
For 6 GHz Model: 100000 Hz to 6000000000 Hz
Default Value: 800000000 Hz
Default Unit: Hz
Example: To set the center frequency to 145 MHz:
:FREQuency:CENTer 145000000
Front Panel Access: Frequency, Rx Freq
[:SENSe]:FREQuency:COUPling OFF|ON|0|1
[:SENSe]:FREQuency:COUPling?
Description: Turns on frequency coupling. When frequency coupling is on, the
Tx frequency cannot be set directly. The Rx Frequency and coupling
offset must be used to set the desired Tx frequency. The Tx frequency
will automatically trail the Rx frequency by the frequency coupling
offset every time the Rx frequency is set. Please note that turning on
frequency coupling will automatically move the Tx frequency to the
Rx frequency plus any frequency coupling offset. If the Rx frequency
and frequency coupling offset is at a setting where the Tx frequency will
be beyond the min/max limits, the instrument will not allow coupling to
be turned on. The query command returns the state of the frequency
coupling setting. A return value of 1 is ON, and a return value of 0 is
OFF.
Cmd Parameters: OFF|ON|0|1
Query Parameters: NA
Range: OFF|ON|0|1
Default Value: OFF
Default Unit: NA
Example: To turn Rx/Tx frequency coupling on:
:SENSe:FREQuency:COUPling ON
Front Panel Access: Frequency, Rx/Tx Coupling