Spectrum Analyzer Commands 4-2 :CALCulate Subsystem
S412E PM PN: 10580-00319 Rev. N 4-5
:CALCulate:LIMit:LOWer:ALARm OFF|ON|0|1
:CALCulate:LIMit:LOWer:ALARm?
Title: Lower Limit Alarm
Description: Enables/disables the lower limit alarm. It is a combination of the
commands :CALCulate:LIMit:TYPe 1 and :CALCulate:LIMit:ALARm
ON|OFF. Setting the value to ON or 1 will turn on the lower limit
alarm. Setting the value to OFF or 0 will turn off the lower limit alarm.
The query version of the command returns a 1 if the lower limit line
alarm is ON and returns 0 if OFF. Note that using this command set
the lower limit line active for editing.
Parameter: OFF|ON|0|1
Parameter Type: <boolean>
Default Value: OFF
Related Command: :CALCulate:LIMit:ALARm
Front Panel Access: Shift-6 (Limit), Limit Alarm
:CALCulate:LIMit:LOWer:FAIL?
Title: Lower Limit Fail State
Description: Query whether the lower limit line has failed or not. The command
returns a 0 on success, 1 on fail, and 2 if the lower limit line is OFF or
the alarm is OFF.
:CALCulate:LIMit:LOWer:POINt?
Title: Number of Lower Limit Points
Description: Returns the number of points currently in the lower limit line.
Default Value: 2
Related Command: :CALCulate:LIMit:POINt?