14-9 :READ Subsystem DMR 2 Commands
14-24 PN: 10580-00319 Rev. N S412E PM
:READ:SIGAnalyzer?
Description: Triggers a new DMR 2 Analyzer measurement and returns the
numerical results. It is a combination of the commands :ABORT;
:INITiate; :FETCh:SIGAnalyzer?
DMR 2 Analyzer must be the active measurement (specified by
:CONFigure:SIGAnalyzer). The current measurement can be queried
using :CONFigure?
For firmware v1.13 and prior, data is returned as 14 comma-separated
values in the following order and format:
Received Power (dBm as float or Watts as long long int or Volts as
long long int)
Frequency Error (Hz as float)
Mod Fid (% as float)
BER (% as float)
Symbol Dev (Hz as float)
CC (hex)
Symbol Rate Error (Hz as float)
Source Address (as integer)
FLCO (as integer)
CSBKO (as integer)
FID (as integer)
Target ID (as integer)
Talk group ID (as integer)
Base Station ID (as integer)
For firmware v1.14 and later, data is returned as 13 comma-separated
values in the following order and format:
Received Power (dBm as float or Watts as long long int or Volts as
long long int)
Frequency Error (Hz as float)
Mod Fid (% as float)
BER (% as float)
Symbol Dev (Hz as float)
CC (hex)
Symbol Rate Error (Hz as float)
Source UID (as integer)
Target ID (as integer)
Talk Group ID (as integer)
FID (as integer)
Call Type (as char)
Base Station ID (as integer)
If there is no valid measurement data, the instrument will return “--” in
that field.