PTC Commands 15-13 [:SENSe] Subsystem
S412E PM PN: 10580-00319 Rev. N 15-41
[:SENSe]:DM:FORMat 4FSK|DQPSK
[:SENSe]:DM:FORMat?
Description: Sets the modulation type. The query will return 4FSK for 4FSK and
DQPS for DSPSK. Please note that setting the modulation type will
restart the sweep.
Cmd Parameters: 4FSK|DQPSK
Query Parameters: NA
Range: 4FSK|DQPSK
Default Value: 4FSK
Default Unit: NA
Example: To set the modulation type to DQPSK:
:DM:FORMat DQPSK
Front Panel Access: Setup, Mod Type
[:SENSe]:DM:SQUelch <value>
[:SENSe]:DM:SQUelch?
Description: Sets the squelch power level. The squelch is only applied to the PTC
Analyzer Summary window on the front panel and will blank out (--) all
summary measurements except for Received Pwr when the received
power level is lower than the squelch power setting.
FETCh:SIGAnalyzer?, READ:SIGAnalyzer?, and
MEASure:SIGAnalyzer? will always return all numerical values.
The query will be returned in the units (dBm, Watts, or Volts) selected
through the Rx Units button using the front panel or with the
command: UNIT:POWer:RX. If the Rx Units has been set to dBm, the
squelch setting is returned in dBm. If the unit is set to Watts, the
squelch setting is returned in fW. If the unit is set to Volts, the squelch
setting is returned in fV.