15-13 [:SENSe] Subsystem PTC Commands
15-42 PN: 10580-00319 Rev. N S412E PM
The set command is sent using the units selected with the Rx Units
button on the front panel or with the command: UNIT:POWer:RX. If the
Rx Units has been set to dBm, the squelch setting must be set in dBm.
If the unit is set to watts, the squelch setting must be set in fW. If the
unit is set to Volts, the squelch setting must be set in fV.
Cmd Parameters: <value>
Query Parameters: NA
Range: 0 dBm to -120 dBm or 1 fW to 1000000000000 fW
or 223.6 mV to 223.61 nV
Default Value: -100 dBm or 100 fW or 2.24 μV
Default Unit: dBm or fW or fV
Example: To set the squelch to -10 dBm:
:DM:SQUelch -10
Front Panel Access: Setup, Squelch Lvl
[:SENSe]:DM:SYMBol:RATE 6000|8000|12000|16000|18000
[:SENSe]:DM:SYMBol:RATE?
Description: Sets the symbol rate of the signal generator.
Cmd Parameters: 6000|8000|12000|16000|18000
Query Parameters: NA
Range: 6000|8000|12000|16000|18000
Default Value: 6000
Default Unit: NA
Example: To set the symbol rate to 16000:
:DM:SYMBol 16000
Front Panel Access: Setup, Symbol Rate