Spectrum Analyzer Commands 4-2 :CALCulate Subsystem
S412E PM PN: 10580-00319 Rev. N 4-15
:CALCulate:MARKer{1|2|3|4|5|6}:DELTa:X <x-parameter>
:CALCulate:MARKer{1|2|3|4|5|6}:DELTa:X?
Title: Delta Marker X Value
Description: Sets the location of the delta marker on the x-axis at the specified
location <x-parameter> + the reference marker x-axis. <x-parameter> is
defined in the current x-axis units. The query version of the command
returns the location of the delta marker on the x-axis.
Parameter: <x-parameter>
Default Unit: Hz or seconds if in zero span
Example: If both the reference and delta marker #1 is currently at 1 GHz on the
x-axis, send the command below to set the delta marker #1 to 2 GHz on
the x-axis:
:CALCulate:MARKer1:DELTa:X 1GHz
(In zero span) If both the reference and delta marker #1 is currently at
25 us on the x-axis, send the command below to set the delta marker to
50us on the x-axis:
:CALCulate:MARKer1:DELTa:X 25us
Related Command: :CALCulate:MARKer[1]|2|3|4|5|6:X
Front Panel Access: Marker, Delta
:CALCulate:MARKer{1|2|3|4|5|6}:DELTa:Y?
Title: Delta Marker Read Y Value
Description: Reads the current Y value for the specified delta marker. The units are
the units of the y-axis.
Default Unit: Current y-axis unit
:CALCulate:MARKer{1|2|3|4|5|6}:DELTa[:SET]:SPAN
Title: Delta Marker to Span
Description: Sets the total span width to the value of the specified delta marker.
Note that this command is valid only if delta marker is on.
Example: To set the span to the value of delta marker #4:
:CALCulate:MARKer4:DELTa:SPAN
Front Panel Access: Marker, More Peak Options, Delta Marker to Span