16-13 [:SENSe] Subsystem NBFM Commands
16-46 PN: 10580-00319 Rev. N S412E PM
[:SENSe]:FILTer:FREQuency 0.3|3|15|NONE
[:SENSe]:FILTer:FREQuency?
Description: Sets the low pass filter frequency.
Note: Low pass filter frequency value is set and returned in kHz
(except for NONE).
Cmd Parameters: 0.3|3|15|NONE
Query Parameters: NA
Range: 0.3|3|15|NONE
Default Value: 3
Default Unit: kHz
Example: To set the low pass filter frequency to 15 kHz:
:FILTer:FREQuency 15
Front Panel Access: Setup, Filter, Low Pass Filter
[:SENSe]:FILTer:HPASs:FREQuency 0.3|3|NONE
[:SENSe]:FILTer:HPASs:FREQuency?
Description: Sets the high pass filter frequency.
Note: High pass filter frequency value is set and returned in kHz
(except for NONE).
Cmd Parameters: 0.3|3|NONE
Query Parameters: NA
Range: 0.3|3|NONE
Default Value: None
Default Unit: kHz
Example: To set the high pass filter frequency to 300 Hz:
:FILTer:HPASs:FREQuency 0.3
Front Panel Access: Setup, Filter, High Pass Filter