4-2 :CALCulate Subsystem Spectrum Analyzer Commands
4-18 PN: 10580-00319 Rev. N S412E PM
:CALCulate:MARKer{1|2|3|4|5|6}:MAXimum:NEXT
Title: Marker (Maximum) Next
Description: Moves the specified marker to the next highest peak anywhere in the
trace that is lower than the current marker. If the specified marker is
not on, the command turns the marker on and sets it to the second
highest peak in the trace. The command uses the existing peak
threshold values to determine the peak value.
Related Command: :CALCulate:MARKer{1|2|3|4|5|6}:MAXimum:LEFT
:CALCulate:MARKer{1|2|3|4|5|6}:MAXimum:RIGHt
Front Panel Access: N/A
:CALCulate:MARKer{1|2|3|4|5|6}:NOISe[:STATe] OFF|ON|0|1
:CALCulate:MARKer{1|2|3|4|5|6}:NOISe[:STATe]?
Title: Marker Noise
Description: Turns the marker noise on or off for the specified reference marker.
Note that if counter marker is set to on when setting marker noise to on,
then counter marker is set to off. This command is not valid in zero
span.
Parameter: OFF|ON|0|1
Parameter Type: <boolean>
Default Value: OFF
Example: To set reference marker #3 as noise marker:
:CALCulate:MARKer3:NOISe ON
:CALCulate:MARKer3:NOISe 1
Front Panel Access: Marker, More, Marker Noise