NBFM Commands 16-13 [:SENSe] Subsystem
S412E PM PN: 10580-00319 Rev. N 16-51
[:SENSe]:OBWidth:PERCent <percentage>
[:SENSe]:OBWidth:PERCent?
Description: This command sets the percentage of carrier power that is used to
measure the occupied bandwidth. This value is used in the
measurement if :SENSe:OBWidth:METHod is set to PERCent.
Cmd Parameters: <percentage>
Query Parameters: NA
Range: 0% to 100%
Default Value: 99%
Default Unit: %
Example: To change the occ bw percent to 70%:
:OBWidth:PERCent 70.00
Front Panel Access: Measurement, NBFM Analyzer, Occ BW Setup, % Int Pwr
[:SENSe]:OBWidth:XDB <rel ampl>
[:SENSe]:OBWidth:XDB?
Description: This command sets the number of dB below the carrier used to measure
the occupied bandwidth. This value is used in the measurement if
:SENSe:OBWidth:METHod is set to XDB.
Cmd Parameters: <rel ampl>
Query Parameters: NA
Range: 0 dBc to 100 dBc
Default Value: 3 dBc
Default Unit: dBc
Example: To change the occ bw XDB to 10:
:OBWidth:XDB 10
Front Panel Access: Measurement, NBFM Analyzer, Occ BW Setup, > dBc