NBFM Commands 16-13 [:SENSe] Subsystem
S412E PM PN: 10580-00319 Rev. N 16-53
[:SENSe:]SWEep:TIME:AUDio <time>
[:SENSe:]SWEep:TIME:AUDio?
Description: Sets the audio sweep time. The setting only applies to the NBFM
Analyzer Audio Waveform graph.
Cmd Parameters: <time>
Query Parameters: NA
Range: 50 us to 150 ms
Default Value: 100 ms
Default Unit: us
Example: To set the audio sweep time to 75 ms:
:SWEep:TIME:AUDio 75000
Front Panel Access: Measurement, NBFM Analyzer, Audio Sweep Time