17-13 [:SENSe] Subsystem TETRA Commands
17-36 PN: 10580-00319 Rev. N S412E PM
[:SENSe]:FREQuency:CENTer <value>
[:SENSe]:FREQuency:CENTer?
Description: Sets the receiver center frequency. Please note that setting the center
frequency will restart the sweep
Cmd Parameters: <value>
Query Parameters: NA
Range: For 1.6 GHz Model: 100000 Hz to 1600000000 Hz
For 6 GHz Model: 100000 Hz to 6000000000 Hz
Default Value: 800000000 Hz
Default Unit: Hz
Example: To set the center frequency to 145 MHz:
:FREQuency:CENTer 145000000
Front Panel Access: Frequency, Rx Freq
[:SENSe]:FREQuency:SPAN 25|50|100|500|1000|5000
[:SENSe]:FREQuency:SPAN?
Description: Sets the span of the Spectrum display in TETRA Analyzer
measurement mode.
Note: Span value is set and returned in kHz.
Cmd Parameters: 25|50|100|500|1000|5000
Query Parameters: NA
Range: 25|50|100|500|1000|5000
Default Value: 25
Default Unit: kHz
Example: To set the span to 1 MHz:
:SENSe:FREQuency:SPAN 1000
Front Panel Access: Frequency, Span