TETRA Commands 17-13 [:SENSe] Subsystem
S412E PM PN: 10580-00319 Rev. N 17-37
[:SENSe]:POWer[:RF]:RANGe[:IMMediate]
Description: Turns off auto ranging and adjusts the receiver reference level once.
In TETRA Analyzer measurement, this command adjusts the receiver
reference level of the spectrum graph.
Cmd Parameters: NA
Query Parameters: NA
Range: NA
Default Value: NA
Default Unit: NA
Example: To adjust range:
:POWer:RANGe
Front Panel Access: Amplitude, Adjust Rx Range
[:SENSe]:POWer[:RF]:RANGe:AUTO OFF|ON|0|1
[:SENSe]:POWer[:RF]:RANGe:AUTO?
Description: Turns auto range for the receiver on or off. When auto range is on, the
reference level is automatically adjusted to the proper value to show the
trace on the screen. If the auto ranging is turned off, the reference level
will not adjust according to where the trace is. In TETRA Analyzer
measurement, this command adjusts the reference level of the spectrum
graph.
Cmd Parameters: OFF|ON|0|1
Query Parameters: NA
Range: OFF|ON|0|1
Default Value: ON or 1
Default Unit: NA
Example: To turn auto ranging off:
:POWer:RANGe:AUTO OFF
Front Panel Access: Amplitude, Auto Rx Range