EasyManua.ls Logo

Anritsu LMR Master S412E - Page 659

Anritsu LMR Master S412E
704 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
AM/FM/PM Commands 18-7 [:SENSe] Subsystem
S412E PM PN: 10580-00319 Rev. N 18-17
[:SENSe]:FREQuency:SIGStandard:NAMe <string>
[:SENSe]:FREQuency:SIGStandard:NAMe?
Title: Signal Standard
Description: Selects the desired signal standard from the list. The <string>
argument is the name of the desired signal standard as displayed in the
instrument’s current signal standard list. The list can be displayed on
the instrument by choosing the Signal Standard submenu button in the
Freq menu. The list can also be downloaded remotely and viewed using
Anritsu Master Software Tools. For example, if the desired Signal
Standard is:
P-GSM 900 - Uplink then the value of the <string> would be “P-GSM
900 - Uplink”.
The query form of this command will return the name of the currently
selected Signal Standard on the list.
Parameter: <string>
Front Panel Access: Freq, Signal Standard
[:SENSe]:FREQuency:SPAN <freq>
[:SENSe]:FREQuency:SPAN?
Title: Frequency Span
Description: Sets the frequency span. Minimum value and the maximum value are
10 kHz and 10 MHz respectively. Note that changing the value of the
frequency span may change the Center Frequency.
Parameter: <freq>
Default Unit: Hz
Front Panel Access: Freq, Span
[:SENSe]:FREQuency:SPAN:FULL
Title: Frequency Span – Full
Description: Sets the frequency span to maximum span (10 MHz). Note that
changing the value of the frequency span may change the Center
Frequency.
Front Panel Access: Freq, Span, Max Span
[:SENSe]:FREQuency:SPAN:MINimum
Title: Frequency Span – Minimum
Description: Sets the frequency span to minimum span (10 kHz). Note that changing
the value of the frequency span may change the Center Frequency.
Front Panel Access: Freq, Span, Min Span