4-14 [:SENSe] Subsystem Spectrum Analyzer Commands
4-54 PN: 10580-00319 Rev. N S412E PM
[:SENSe]:AVERage:TYPE NONE|SCALar|MAXimum|MINimum
[:SENSe]:AVERage:TYPE?
Title: Trace Mode (Normal/Average/Max Hold/Min Hold)
Description: Specifies how successive traces are combined to produce the resulting
display value. Setting the TYPE to NONE is the equivalent of setting
the trace mode to “Normal->A” on the front panel. The displayed value
for a point is the current measured value for that point. Setting the
TYPE to SCALar is the equivalent of setting the trace mode to
“Average->A” on the front panel. The displayed value for a point is the
average of the last <integer> measured values where <integer> is set by
[:SENSe]:AVERage:COUNt. Setting the TYPE to MAXimum is the
equivalent of setting the trace mode to “Max Hold->A” on the front
panel. The displayed value for a point is the maximum measured value
for that point over sweeps. Setting the TYPE to MINimum is the
equivalent of setting the trace mode to “Min Hold->A” on the front
panel. The displayed value for a point is the minimum measured value
for that point over sweeps.
Parameter: NONE|SCALar|MAXimum|MINimum
Parameter Type: <char>
Default Value: NONE
Example: To set the TYPE to SCALar:
:SENSe:AVERage:TYPE SCALar
To set the TYPE to MAXimum:
:SENSe:AVERage:TYPE MAXimum
Related Command: :AVERage:COUNt
Front Panel Access: Shift-5 (Trace), Trace A Operations
[:SENSe]:BANDwidth|BWIDth:VIDeo <freq>
[:SENSe]:BANDwidth|BWIDth:VIDeo?
Title: Video Bandwidth
Description: Sets the video bandwidth. Note that using this command turns the
automatic video bandwidth setting OFF.
Parameter: <freq>
Default Value: 1 MHz
Default Unit: Hz
Range: 1 Hz to 3 MHz in a 1:3 sequence except for the MS2711E
10 Hz to 3 MHz in a 1:3 sequence for the MS2711E
Related Command: :BANDwidth|BWIDth:VIDeo:AUTO
Front Panel Access: BW, VBW