Performance Verification 2 3-9 WCDMA/HSDPA Signal Analyzer (Options 44, 45, and 65) Verification
MS2721B MM PN: 10580-00177 Rev. D 3-49
37. On the MG3700A, turn on the RF output and use the knob to adjust power level to read the value of
PMA.10C on Sensor A refer to Table A-100).
38. Record the MG3700A power level setting (MG3700A.10) in Table A-101, “Power Level Settings
Component Characterization, Option 44” on page A-34.
39. On the MS2721B, press the Amplitude function hard key and then press the Adjust Range soft key.
40. Record the channel power reading in the "Measured Power" column of Table A-102, “WCDMA Absolute
Power Accuracy, Option 44” on page A-34.
41. Use the following formula to calculate the absolute power accuracy of the MS2721B at +10 dBm:
Error = Measured Power 10
42. Record the calculated value in the "Error" column of Table A-102, and verify that it is within
specification.
43. Turn off the RF output of the MG3700A.
44. Set the MN63A attenuator to 18 dB.
45. Calculate the value of the MG3700A setting (MG3700A.28) for +28 dBm Test Level by using Table A-100
and Table A-101 and the following formula:
MG3700A.28 = MG3700A.10 + ATT.18
46. Record the calculated value in Table A-101.
47. On the MG3700A, turn on the RF output and use the knob to adjust power level to the recorded
MG3700A.28 value in Table A-101.
48. On the MS2721B, press the Amplitude function hard key and then press the Adjust Range soft key.
49. Record the channel power reading in the "Measured Power" column of Table A-102.
50. Use the following formula to calculate the absolute power accuracy of the MS2721B at +28 dBm:
Error = Measured Power 28
51. Record the calculated value in the "Error" column of Table A-102, and verify that it is within
specification.
52. Turn off the RF output of the MG3700A.
53. Set the MN63A attenuator to 0 dB.
54. Set power level of the MG3700A to 38 dBm.
55. Calculate the value of the Sensor A reading (PMA10) for –10 dBm Test Level by using the following
formula:
PMA-10 = PMA.10C 20
56. Record the calculated value in Table A-101.
57. Turn on the RF output and use the knob to adjust power level to read the value of PMA10 on Sensor A.
58. On the MS2721B, press the Amplitude key and then press the
Adjust Range soft key.
59. Record channel power reading in Table A-102.
60. Use the following formula to calculate the absolute power accuracy of the MS2721B at 10 dBm:
Error = Measured Power (–10)
61. Record the calculated Error value in Table A-102, and verify that it is within specification.
62. Turn off the RF output of the MG3700A.
63. Set the MN63A attenuator to 0 dB.
64. Decrease power level of the MG3700A by 10 dB.

Table of Contents

Other manuals for Anritsu Spectrum Master MS2721B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Anritsu Spectrum Master MS2721B and is the answer not in the manual?

Anritsu Spectrum Master MS2721B Specifications

General IconGeneral
BrandAnritsu
ModelSpectrum Master MS2721B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals