Performance Verification 2 3-15 Mobile WiMAX Residual EVM and Frequency Error Tests (Option 67)
MS2721B MM PN: 10580-00177 Rev. D 3-103
Residual EVM and Frequency Error (10 MHz Bandwidth and 10 ms Frame Length)
3. Set the MG3700A as follows:
a. Press the yellow Preset button (answer yes to the question).
b. Press the Set key (Note that two Set keys are available, and they both do the same thing).
c. Press the (F1) soft key to select "Load File to Memory"
d. Press the (F1) soft key again to select "Select Package"
e. Using the Down Arrow key, step through the selection list until "mWiMax" is highlighted.
f. Press the Set key.
g. Press the (F6) soft key, "Return"
h. Press the Set key. The "Select Package" list box appears. Again select "mWiMax" and press Set.
i. Another file list appears. Select (highlight) "10m1024g8_0_10_cap".
j. Press the Set key.
k. Press the MOD On/Off key and verify that the LED is ON (illuminated). Ensure that the "playing"
indicator is displaying the moving pattern.
l. Press the Frequency key, then enter 2600.5 MHz.
m. Press the Level key, then enter 15 dBm. Turn on the output.
4. Set the MS2721B to Mobile WiMax Signal Analyzer Mode and preset the instrument.
5. Set the MS2721B as follows:
a. Press the Freq function hard key and set the Center Freq to 2600.5 MHz
b. Press the Setup function hard key and set the Bandwidth to 10 MHz.
c. Press the CP Ratio G soft key (in the Setup menu) and set the CP Ratio to 1/8.
d. Press the Frame Length soft key and set the Frame Length to 10 ms.
e. Press the Demod soft key and set Demod to FCH.
f. Press the Measurements function hard key and press Demodulator, then press
Modulation Summary.
6. Record the MS2721B EVM (rms) reading in Table A-152, “Mobile WiMAX Residual EVM (10 MHz
Bandwidth and 10 ms Frame Length), Option 67” on page A-49.
7. Verify that the measured EVM is within specification.
8. Set the MG3700A Level to 50.0 dBm.
9. Record the MS2721B EVM (rms) reading in Table A-152.
10. Verify that the measured EVM is within specification.
11. On the MS2721B, press the Setup function hard key and set Demod to Auto.
12. Record the MS2721B Frequency Error reading in Table A-154, “Mobile WiMAX Frequency Error
(10 MHz Bandwidth and 10 ms Frame Length), Option 67” on page A-49.
13. Verify that the measured Frequency Error is within specification.
14. Set MG3700A Frequency to 3600.5 MHz.
15. On the MS2721B, change the Center Frequency to 3600.5 MHz.
16. Press the Setup function hard key and set Demod to FCH.
17. Measure the EVM (rms) for –50 dBm and then record the measured results in Table A-152.
18. Verify that the measured EVM is within specification.
19. On the MS2721B, press the Setup function hard key and set Demod to Auto.

Table of Contents

Other manuals for Anritsu Spectrum Master MS2721B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Anritsu Spectrum Master MS2721B and is the answer not in the manual?

Anritsu Spectrum Master MS2721B Specifications

General IconGeneral
BrandAnritsu
ModelSpectrum Master MS2721B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals