3-4 ISDB-T Signal Analyzer (Option 30) Verification Performance Verification 2
3-10 PN: 10580-00177 Rev. D MS2721B MM
30. Wait until Average (10/10) appears at the top of the display.
31. Record the Frequency Error in Table A-45.
Level Accuracy Verification
The tests in this section verify the level accuracy of the MS2721B in ISDB-T Signal Analyzer mode.
Required Equipment:
Anritsu MG3700A Vector Signal Generator
Programmable Attenuator, Anritsu Model MN63A
Power Meter, Anritsu Model ML2438A
Power Sensor, Anritsu Model MA2482D (2)
Anritsu 1N50C RF Limiter
Adapter, Anritsu Model 34NN50A
RF Coaxial Cable, Anritsu Model 15NN50-1.5C
Power Splitter, AeroflexWeinschel Model 1870A
Fixed Attenuator, Aeroflex/Weinschel Model 44-10
RF Power Amplifier, Mini Circuits Model TIA-1000-1R8
Midwest Microwave ADT-2615-NF-BNM-02 Adapters (2)