Performance Verification 2 3-9 WCDMA/HSDPA Signal Analyzer (Options 44, 45, and 65) Verification
MS2721B MM PN: 10580-00177 Rev. D 3-45
3-9 WCDMA/HSDPA Signal Analyzer (Options 44, 45, and 65) Verification
The tests in this section can be used to verify the functionality of the WCDMA/HSDPA Signal Analyzer mode of
the MS2721B Spectrum Master. The tests include:
WCDMA Absolute Power Accuracy Verification (Option 44)
WCDMA Occupied Bandwidth (OBW) Verification (Option 44)
WCDMA RF Channel Power Accuracy and ACLR Verification (Option 44)
HSDPA RF Channel Power Accuracy and ACLR Verification (Option 44)
Error Vector Magnitude (EVM) Verification (Options 45 and 65)
WCDMA Absolute Power Accuracy Verification (Option 44)
This test verifies the WCDMA absolute power accuracy in WCDMA/HSDPA Signal Analyzer Mode.
Required Equipment:
Anritsu MG3700A Vector Signal Generator
RF Power Amplifier, Mini Circuits Model TIA-1000-1R8
Frequency Reference Symmetricom Rubisource T&M
Power Meter, Anritsu Model ML2438A
Programmable Attenuator, Anritsu Model MN63A
Circulator
High Power Load, AeroflexWeinschel Model M1418
Coupler
Power Sensor, Anritsu Model MA2482D (2)
Adapter, Anritsu Model 34NN50A (2)
Spectrum Master Model MS2721B

Table of Contents

Other manuals for Anritsu Spectrum Master MS2721B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Anritsu Spectrum Master MS2721B and is the answer not in the manual?

Anritsu Spectrum Master MS2721B Specifications

General IconGeneral
BrandAnritsu
ModelSpectrum Master MS2721B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals