3-3 Tracking Generator (Option 20) Verification Performance Verification 2
3-4 PN: 10580-00177 Rev. D MS2721B MM
12. Change the MS2721B Center Frequency to match the Frequency in the next test record in Appendix A,
which is Table A-39, “Level Accuracy 975 MHz – Tracking Generator (Option 20)” on page A-14.
13. Set the power sensor calibration factor frequency to the same frequency.
14. Repeat the measurements for all frequencies and power levels shown in the test records of Table A-39
through Table A-42, “Level Accuracy 7075 MHz – Tracking Generator (Option 20)” on page A-14.