Performance Verification 2 3-4 ISDB-T Signal Analyzer (Option 30) Verification
MS2721B MM PN: 10580-00177 Rev. D 3-21
8. Press the Meas Setup function hard key and then the Meas Mode soft key. Use the Down Arrow key to
select Average, and then press the Enter key.
9. Wait until the Average counter displays (10/10).
10. Record the 10 kHz and the 100 kHz Phase Noise readouts into the “Phase (10 kHz)” and the
“Phase (100 kHz)” columns in the “Channel 13” row in Table A-60, “ISDB-T Signal Analyzer Phase
Noise, Option 30” on page A-21.
11. Record the Frequency Error in Table A-60.
12. Set the Frequency of the MG3700A to 623.14285714 MHz and change the MS2721B Channel to 38.
13. Wait until the Average counter displays (10/10).
14. Record the 10 kHz and the 100 kHz Phase Noise readouts in the “Channel 38” row in Table A-60.
15. Record the Frequency Error in Table A-60.
16. Set the Frequency of the MG3700A to 767.14285714 MHz and change the MS2721B Channel to 62.
17. Wait until Average counter displays (10/10).
18. Record the 10 kHz and the 100 kHz Phase Noise readouts in the “Channel 62” row in Table A-60.
19. Record the Frequency Error in Table A-60.

Table of Contents

Other manuals for Anritsu Spectrum Master MS2721B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Anritsu Spectrum Master MS2721B and is the answer not in the manual?

Anritsu Spectrum Master MS2721B Specifications

General IconGeneral
BrandAnritsu
ModelSpectrum Master MS2721B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals