Chapter 7 Verifying the Instrument Performance
Preparing the TaqMan
®
RNase P Verification Plate
90 Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide
Notes
Preparing the TaqMan
®
RNase P Verification Plate
IMPORTANT! You must perform an ROI calibration
(see page 35), a background calibration (see page 49)
and a pure dye calibration (see page 67) before
running an RNase P plate.
IMPORTANT! Wear powder-free
gloves when you handle the
RNase P
Verification plate
.
1. Retrieve the TaqMan RNase P Verification Plate
Kit from the freezer, remove the RNase P plate,
then allow the plate to warm to room temperature
(approximately 5 minutes).
2. Remove the
RNase P
plate from its packaging.
Note: If the
RNase P
plate
contains a
compression pad, remove it from the plate.
Applied Biosystems does not recommend the use
of compression pads with the 7300/7500/7500
Fast system.
3. Standard plates only: Vortex the plate for
5 seconds. Do not vortex Fast plates.
(Remaining steps apply to both standard and Fast
plates.)
4. Centrifuge the plate for 2 minutes in a centrifuge
with a plate adapter (<1500xg).
IMPORTANT! The plate must be well mixed and
centrifuged.
>00:05:00Ambient
RNase P
Verification Plate
00:00:05<1500× g