Chapter 7 Verifying the Instrument Performance
Preparing the TaqMan
®
RNase P Verification Plate
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide 91
Notes
7
5. Verify that the liquid in each well of the plate is
at the bottom of the well. If not, centrifuge the
plate again at a higher speed and for a longer
period of time.
Continue with “Creating a Plate Document for the
RNase P Verification Run” on page 92.
Correct Incorrect
Liquid is at
bottom of well.
• Not centrifuged with enough force,
or
• Not centrifuged for enough time