Appendix A
Replacing the Instrument Fuses
122 Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide
Notes
Replacing the Instrument Fuses
FIRE HAZARD. For continued protection against the risk of fire,
replace fuses only with listed and certified fuses of the same type and rating as those
currently in the instrument.
Time Required 30 minutes
Materials
Required
Replacing the
Fuses
1. Turn off the instrument, then unplug it.
2. Using a flat-head screwdriver, unscrew and remove the fuse holders from the
instrument.
Powder-free
gloves
Flathead
screwdriver
Fuses (2), 12.5A,
250V, 5 x 20 mm
Safety
glasses