EasyManua.ls Logo

Applied Biosystems 7300 - Troubleshooting - ROI Image

Applied Biosystems 7300
174 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 4 Performing the Regions of Interest (ROI) Calibration
Troubleshooting
48 Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide
Notes
Troubleshooting
Troubleshooting – ROI Image
Condition: The ROI image is faint or you cannot
generate a successful calibration.
The sample block may be in its lowered position.
If the CCD Control Settings in the ROI Inspector
displays , click to raise
the sample block.
Check that the heated cover assembly is pulled all
the way forward to ensure that the tray can be
pushed in properly.
Click
GR2482
7500/7500 fast
heated cover assembly
GR2482
Heated cover
assembly

Table of Contents

Other manuals for Applied Biosystems 7300

Related product manuals